![Keysight 53210A User Manual Download Page 223](http://html1.mh-extra.com/html/keysight/53210a/53210a_user-manual_1961097223.webp)
Instrument Status
8
Keysight 53210A User’s Guide
223
Keysight 53210A Status System
This chapter provides an overview of the 53210A Status System. Refer to the
STATus
subsystem and the IEEE-488 commands within the Programmer’s
Reference for bit definitions and additional information. The reference is located
on the 53210A/53220A/53230A Product Reference CD (p/n 53220-13601).
Questionable Data Register Group
The counter’s Questionable Data Register Group monitors the following
conditions and events within the counter:
– input voltage overload
– frequency measurement hardware time out
– calibration data
– time base error
– lower-limit exceeded
– upper limit exceeded
– reading memory overflow
Condition Register
The Condition Register within the Questionable Data Register Group monitors
current conditions that may impact the integrity of the counter measurements.
The Condition Register is read with the command:
STATus:QUEStionable:CONDition?
Reading the register does not clear the bit(s) in the register. The bit is cleared
when the condition is removed.
Event Register
The Event Register within the Questionable Data Register Group also monitors
conditions that may impact the integrity of the counter measurements.
Summary of Contents for 53210A
Page 1: ...Keysight 53210A 350 MHz RF Counter User s Guide ...
Page 12: ...12 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...22 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 246: ...246 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...