l
When data logging to file(s),
Browse
to an internal or external path and specify a
File Name
. If more
than one file needs to be created to hold the data, the second file name will be appended with _00001,
the third file name with _00002, and so on. When data logging to files, the maximum number of read-
ings is 360,000,000.
When
Add Date
is
On
, the data logging start date and time is appended to the file name using the
format:
_YYYYMMDD_HHMMSS
For example, for a file named
Data 1
, the result will be similar to:
Data 1_20140720_032542
.
Press
Options
to to configure reading storage options:
Rows/File
- Specifies the maximum number of rows or readings to be written to a file. For
Max
, the
limit is the number of bytes allowed by the file system (2
32
= 4.294967296 GBytes). This represents
approximately 252 M readings with Metadata Off, or 159 M readings with Metadata On. For
1M
, the
limit is 1,000,000 rows in the resulting file. This allows you to accommodate common spreadsheet,
database and data analysis programs that have limitations of 1 million rows per file.
Metadata
- Enables reading number, time stamp of first reading, and sample interval (if available) in
the file.
Separator
- Specifies the character (Comma, Tab, or Semicolon) to use for separating the information
on each row.
When finished configuring reading storage, press
Done
>
Done
to return to the main data logging
menu.