AC Current Gain and Flatness Calibration
This calibration procedure applies to the 34460A and 34461A only.
Review
and
Flatness Calibration Considerations
before performing any calibration.
Configuration: AC Current
All AC calibration procedures use the 3 Hz bandwidth measurement filter.
1.
Select each range in the order shown in the table below.
2.
Apply each input current and frequency shown in the Input columns.
3.
Enter the actual applied input current as the Cal Value (see
). A Calibration
Step Succeeded message indicates success; if the display shows Calibration Step Failed, check the
input value, range, function, and entered calibration value and repeat the calibration step.
4.
Repeat steps 1 through 3 for each frequency shown in the table.
5.
Repeat steps 1 through 4 for each current range in the table.
6.
Store the new calibration constants (see
Storing the Calibration Constants
).
Each current and frequency takes approximately 2.5 seconds to complete.
Input
Instrument
Current
Frequency
Range
1 A
5 kHz, 7.5 kHz, 9.7 kHz, 10 kHz, 220 Hz, 1 kHz
1 A
100 mA 5 kHz, 7.5 kHz, 9.7 kHz, 10 kHz, 220 Hz, 1 kHz
100 mA
10 mA
5 kHz, 7.5 kHz, 9.7 kHz, 10 kHz, 220 Hz, 1 kHz
10 mA
1 mA
1 kHz
1 mA
100 µA
1 kHz
100 µA
Keysight True
volt
Series Operating and Service Guide
545
Calibration Adjustment Procedures