AC Voltage and AC Current Verification
This procedure verifies AC voltage and current.
1.
Make sure you have read
.
2.
Connect the calibrator to the input terminals. For the 34461A/65A/70A, use the front panel input ter-
minals and select the front input terminals with the
Front/Rear
switch.
3.
Set the AC Volts or AC Current function and the filter as shown below.
4.
Select each range in the order shown below. Provide the indicated input voltage and frequency. Com-
pare measurement results to the test limits shown below. (Be certain to allow for appropriate source
settling.)
ACV Values for 34460A DMM
Input
Error from Nominal
Vrms
Frequency Range
Filter
Quick
Check
24 hours 90 days
1 year
2 years
100 mV
1 kHz
100 mV 200 Hz
± 90 µV
± 110 µV ± 120 µV ± 130 µV
50 kHz
± 170 µV ± 190 µV ± 200 µV ± 210 µV
300 kHz
Q
± 4.5 mV
1 V
1 kHz
1 V
200 Hz
± 900 µV ± 1.1 mV ± 1.2 mV ± 1.3 mV
50 kHz
± 1.7 mV ± 1.9 mV
± 2 mV
± 2.1 mV
300 kHz
± 45 mV
0.03 V
1 kHz
10 V
200 Hz
± 2 mV
± 3 mV
1 V
1 kHz
200 Hz
± 2.7 mV ± 3.8 mV ± 3.9 mV
± 4 mV
10 V
10 Hz
3 Hz
± 9 mV
± 11 mV
± 12 mV
± 13 mV
100 Hz
20 Hz
± 9 mV
± 11 mV
± 12 mV
± 13 mV
20 kHz
200 Hz
± 9 mV
± 11 mV
± 12 mV
± 13 mV
50 kHz
200 Hz
Q
± 17 mV
± 19 mV
± 20 mV
± 21 mV
100 kHz
± 66 mV
± 71 mV
300 kHz
± 450 mV
100 V
1 kHz
100 V
200 Hz
Q
± 90 mV ± 110 mV ± 120 mV ± 130 mV
50 kHz
± 170 mV ± 190 mV ± 200 mV ± 210 mV
70 V
300 kHz
± 3.3 V
750 V
1 kHz
750 V
200 Hz
± 675 mV ± 825 mV ± 900 mV ± 975 mV
210 V
50 kHz
± 573 mV ± 669 mV ± 690 mV ± 711 mV
70 V
300 kHz
± 6.6 V
Q
:
Quick performance verification test point.
Performance Verification
516
Keysight True
volt
Series Operating and Service Guide