TRIGger:SLOPe {POSitive|NEGative}
TRIGger:SLOPe?
Selects whether the instrument uses the rising edge (POS) or the falling edge (NEG) of the trigger signal on
the rear-panel
Ext Trig
BNC connector when external triggering is selected; or the rising or falling edge of
the input signal when level triggering is selected (level triggering applies only to the 34465A and 34470A).
Parameter
Typical Return
{POSitive|NEGative}. Default: NEG
POS or NEG
Return ten sets of five DC voltage measurements, using a positive-going external trigger to start each meas-
urement set:
CONF:VOLT:DC
SAMP:COUN 5
TRIG:COUN 10
TRIG:SOUR EXT;SLOP POS
READ?
Typical Response: +1.0001, … (50 measurements)