CONFigure:DIODe
Sets all measurement parameters and trigger parameters to their
for diode tests.
Parameter
Typical Return
(none)
(none)
Configure, make, and read a default diode measurement:
CONF:DIOD
READ?
Typical Response: +1.32130000E-01
l
The range and resolution are
fixed
for diode tests: the range is 1 VDC (with a 1 mA current source out-
put).
l
The voltage is displayed on the front panel if it is between 0 and 5.05 V. The instrument beeps when
the signal transitions into the range between 0.3 and 0.8 V (unless beeper is disabled). If the signal
exceeds 5.05 V, the front panel shows "OPEN", and the value returned from SCPI is 9.9E37.
l
The
, and
queries return the measured voltage, regardless of its
value.
l
Use
or
to start the measurement.