CONFigure:CONTinuity
Sets all measurement parameters and trigger parameters to their
for continuity meas-
urements.
Parameter
Typical Return
(none)
+1.32130000E-02
Configure the instrument for continuity measurements. Then make a measurement using an external trig-
ger with positive slope (rising edge) and read the measurement:
CONF:CONT
TRIG:SOUR EXT;SLOP POS
READ?
l
The range and resolution are fixed at 1 kΩ for continuity tests (a 2-wire resistance measurement).
l
The instrument beeps (if the beeper is enabled) for each measurement less than or equal to the con-
tinuity threshold (≤10 Ω), and the actual resistance measurement appears on the display.
l
From 10 Ω to 1.2 kΩ, the instrument displays the actual resistance measurement with no beep. Above
1.2 kΩ, the instrument displays "OPEN" with no beep.
l
The
, and
queries return the measured resistance, regardless of
its value.
l
Use
or
to start the measurement.
SCPI Programming Reference