
CALibration:SECure:STATe {ON|1|OFF|0}[, <
code
>]
CALibration:SECure:STATe?
Unsecures or secures the instrument for calibration.
You must unsecure the instrument to calibrate it or to change the calibration code or security code.
Parameter
Typical Return
{ON|1|OFF|0}. Default: ON. 0 (OFF) or 1 (ON)
Unsecure calibration:
CAL:SEC:STAT OFF,MY_CODE_272
Secure calibration:
CAL:SEC:STAT ON
l
The <
code
> is optional to secure the instrument, but must be correct if provided.
l
When shipped from the factory, the instrument is secured, with the security code (calibration pass-
word) set to
AT3446XA
.
l
Front panel and remote interface calibration share the same code.
l
This setting is non-volatile; it will not be changed by power cycling or
or
.
CALibration:STORe
Takes calibration constants in volatile memory (
), and places them in nonvolatile
memory, where they are not changed by cycling power,
, or
. Do this at the end of cal-
ibration to avoid losing changes.
You must unsecure the instrument to calibrate it or to change the calibration code or security code.
Parameter
Typical Return
(none)
(none)
Store calibration constants into non-volatile memory:
CAL:STOR
SCPI Programming Reference
240
Keysight True
volt
Series Operating and Service Guide