CALCulate:TRANsform:HISTogram:CLEar[:IMMediate]
Clears the histogram data and restarts histogram ranging if it is enabled (
late:TRANsform:HISTogram:RANGe:AUTO ON
).
Parameter Typical Return
(none)
(none)
Clear the histogram data:
CALC:TRAN:HIST:CLE
l
Does not clear measurements in reading memory.
l
To clear statistics, limits, histogram data, and measurement data, use
.
CALCulate:TRANsform:HISTogram:COUNt?
Returns the number of measurements collected since the last time the histogram was cleared.
Parameter
Typical
Return
(none)
+87
Return the number of measurements used to compute the current histogram:
CALC:TRAN:HIST:COUN?
CALCulate:TRANsform:HISTogram:POINts {<
value
>|MIN|MAX|DEF}
CALCulate:TRANsform:HISTogram:POINts? [{MIN|MAX|DEF}]
Sets the number of bins between the lower and upper range values for the histogram. Two additional bins
always exist: one for measurements below the lower range and one for measurements above the upper
range.
Parameter
Typical Return
{10|20|40|100|200|400|MIN|MAX|DEF}. Default: 100. +100
See
l
You can specify the lower and upper range values using
CALCulate:TRANsform:HISTogram:RANGe:
. Lower and upper range values are computed automatically if
late:TRANsform:HISTogram:RANGe:AUTO
is ON.
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
Keysight True
volt
Series Operating and Service Guide
219
SCPI Programming Reference