Math - Limits
Limit checking indicates how many samples have exceeded specified limits and visually indicates when the
limits are exceeded.
Front Panel Menus
The limits menu is accessed from the
[Math]
menu.
The first softkey enables or disables limits. The second and third softkeys specify the limits either as high
and low values or as a span around a center value. For example, a
Low
limit of -4 V and a
High
limit of +7 V
are equivalent to a
Center
of 1.5 V and a
Span
of 11 V.
The
Beeper
softkey enables or disables beeping when limits are violated (this also enables or disables the
beeper for the other functions that use the beeper - probe hold, diode, continuity, and errors.
Clear Condi-
tion
resets the limit borders to green, as described below.
Features and Functions
110
Keysight True