Chapter 4 Calibration Procedures
0 dB Range Flatness Verification
4
4
0 dB Range Flatness Verification
This procedure checks the high frequency ac amplitude flatness above
100 kHz on the 0dB attenuator range. (Flatness is relative to 1 kHz.)
1
Connect the power meter to measure the output amplitude of the
instrument as shown below.
2
Set up the function generator as follows:
• Output impedance: 50
Ω
(press
and select Output Setup).
• Waveform: Sine
• Frequency: 100 kHz
• Amplitude: 3.51 dBm
Make sure the output is enabled.
3
On the power meter, use the Relative Power function to set the current
reading as the reference value. This will allow you to compare future
measurement results in dB.
4
Set the power meter offset to equal the 100kHz_0dB_offset value
previously calculated. This sets the power meter to directly read the
flatness error specification relative to 1 kHz. 100kHz_0dB_offset is
calculated on the Amplitude and Flatness Verification Worksheet.
Summary of Contents for 33210A
Page 1: ...Keysight 33210A 10 MHz Function Arbitrary Waveform Generator Service Guide ...
Page 2: ......
Page 3: ......
Page 10: ...8 ...
Page 13: ...1 1 Specifications ...
Page 19: ...2 2 Quick Start ...
Page 37: ...3 3 Front Panel Menu Operation ...
Page 52: ...50 3 ...
Page 53: ...4 4 Calibration Procedures ...
Page 96: ...94 Chapter 4 Calibration Procedures Calibration Errors 4 ...
Page 97: ...5 5 Block Diagram ...
Page 101: ...99 Chapter 5 Block Diagram Block Diagram 4 5 ...
Page 103: ...6 6 Disassembly and Repair ...
Page 128: ...126 Chapter 6 Disassembly and Repair Replaceable Parts 6 ...
Page 129: ...7 7 Backdating ...