41
Chapter 3 Front-Panel Menu Operation
To Unsecure and Secure for Calibration
4
3
To Unsecure and Secure for Calibration
This feature allows you to enter a security code to prevent accidental
or unauthorized adjustments of the instrument. When you first receive
your instrument, it is secured. Before you can adjust the instrument,
you must unsecure it by entering the correct security code.
• The security code is set to
AT33210A
when the instrument is shipped
from the factory. The security code is stored in non-volatile memory,
and does not change when power has been off, after a Factory Reset
(
*RST
command), or after an Instrument Preset (
SYSTem:PRESet
command).
• The security code may contain up to 12 alphanumeric characters.
The first character must be a letter, but the remaining characters
can be letters, numbers, or an underscore ( _ ). You do not have to
use all 12 characters but the first character must always be a letter.
Note
If you forget your security code, you can disable the security feature by
applying a temporary short inside the instrument as described in
“To Unsecure the Instrument Without the Security Code” on page 71.
Summary of Contents for 33210A
Page 1: ...Keysight 33210A 10 MHz Function Arbitrary Waveform Generator Service Guide ...
Page 2: ......
Page 3: ......
Page 10: ...8 ...
Page 13: ...1 1 Specifications ...
Page 19: ...2 2 Quick Start ...
Page 37: ...3 3 Front Panel Menu Operation ...
Page 52: ...50 3 ...
Page 53: ...4 4 Calibration Procedures ...
Page 96: ...94 Chapter 4 Calibration Procedures Calibration Errors 4 ...
Page 97: ...5 5 Block Diagram ...
Page 101: ...99 Chapter 5 Block Diagram Block Diagram 4 5 ...
Page 103: ...6 6 Disassembly and Repair ...
Page 128: ...126 Chapter 6 Disassembly and Repair Replaceable Parts 6 ...
Page 129: ...7 7 Backdating ...