107
Chapter 6 Disassembly and Repair
Troubleshooting Hints
4
6
Troubleshooting Hints
This section provides a brief check list of common failures. Before
troubleshooting or repairing the instrument, make sure the failure is
in the instrument rather than any external connections. Also make
sure that the instrument is accurately calibrated within the last year
(see “Calibration Interval”, on page 53). The instrument’s circuits allow
troubleshooting and repair with basic equipment such as a 6
½
digit
multimeter.
Unit is Inoperative
• Verify that the ac power cord is connected to the instrument.
• Verify that the front-panel On/Standby switch has been pushed.
Is the display working?
• If the display appears blank, but front-panel keys are lit (and the fan
is working), the display screen saver may be enabled. Press any key
to restore the display.
• If the display appears blank, but front-panel keys are lit (and the fan
is working), the display contrast setting may be too low. Use the
menu to set the display contrast.
Unit Fails Self-Test
Ensure that all terminal connections (both front panel and rear terminals)
are removed while the self-test is performed.
Summary of Contents for 33210A
Page 1: ...Keysight 33210A 10 MHz Function Arbitrary Waveform Generator Service Guide ...
Page 2: ......
Page 3: ......
Page 10: ...8 ...
Page 13: ...1 1 Specifications ...
Page 19: ...2 2 Quick Start ...
Page 37: ...3 3 Front Panel Menu Operation ...
Page 52: ...50 3 ...
Page 53: ...4 4 Calibration Procedures ...
Page 96: ...94 Chapter 4 Calibration Procedures Calibration Errors 4 ...
Page 97: ...5 5 Block Diagram ...
Page 101: ...99 Chapter 5 Block Diagram Block Diagram 4 5 ...
Page 103: ...6 6 Disassembly and Repair ...
Page 128: ...126 Chapter 6 Disassembly and Repair Replaceable Parts 6 ...
Page 129: ...7 7 Backdating ...