86
Keysight 1000B Series Oscilloscopes User’s Guide
3
Capturing Data
7
Press
Setting
and turn the entry knob to adjust the width setting.
To set up video triggers
Video triggering is used to trigger on fields or lines of NTSC, PAL, or SECAM
standard video waveforms.
When the video trigger mode is selected, the trigger coupling is set to AC.
1
Press
[Menu]
.
2
In the Trigger menu, press
Mode
.
3
Continue pressing the
Mode
softkey or turn the entry knob to select “Video”.
4
Then, either push the entry knob or press
Mode
again.
5
Press
Polarity
to toggle between:
6
Press
Sync
and continue pressing the softkey or turn the entry knob to select
what to trigger on:
7
Press
Standard
to toggle between:
Normal polarity — trigger on the negative edge of the sync
pulse.
Inverted polarity — trigger on the positive edge of the sync
pulse.
NOTE
Normal Polarity Sync triggers always occur on negative-going horizontal sync pulses. If the
video waveform has positive-going horizontal sync pulses, use the Inverted Polarity selection.
All Lines
Trigger on all lines.
Line Num
Trigger on a selected line.
If you select “Line Num”, press the following
Line Num
menu
item and turn the entry knob to select the line number.
Odd Field
Trigger on an odd field.
Even Field
Trigger on an even field.
Summary of Contents for 1000B Series
Page 1: ...Keysight 1000B Series Oscilloscopes User s Guide ...
Page 12: ...12 Keysight 1000B Series Oscilloscopes User s Guide Contents ...
Page 16: ...16 Keysight 1000B Series Oscilloscopes User s Guide Tables ...
Page 36: ...36 Keysight 1000B Series Oscilloscopes User s Guide 1 Getting Started ...
Page 124: ...124 Keysight 1000B Series Oscilloscopes User s Guide 5 Saving Recalling and Printing Data ...
Page 138: ...138 Keysight 1000B Series Oscilloscopes User s Guide 6 Oscilloscope Utility Settings ...