U9361C/F/G/M RCal Receiver Calibrator Measurement Guide
47
Demonstrations
Demo #5 - RCal vs. Power Sensor Accuracy
Compatible Signal Generators:
— N517xB (all EXG)
— N518xB (all MXG)
— N519xA (all UXG)
— M9x84B/C (all VXG)
This demonstration compares the absolute amplitude accuracy after an RCal calibration compared
to a calibration done using a power sensor. In this case, we used the U5532C sensor module, but
this demonstration can be done with other types of sensor modules and power meters.
1.
Preset the analyzer.
Mode
>
Spectrum Analyzer
>
Mode Preset
2.
Connect the signal analyzer to the signal generator.
3.
Set start frequency to 1 GHz and stop frequency to 20 GHz.
FREQ
>
Start Freq
>
1 GHz
Stop Freq
>
20 GHz