U9361C/F/G/M RCal Receiver Calibrator Measurement Guide
11
RCal Software – Configuration
Shown below are all parameters (columns) for an RCal calibration state (row):
RCal Configuration Table – Parameters
The following table details the parameters for each column in every calibration row.
Column Title
Description
Additional Information
Calibrate
(checkbox)
Rows that are selected will be included when
Calibrate Checked Rows
is used to start a
calibration.
Apply
(checkbox)
Rows that are successfully calibrated will be
applied to measurements when selected.
Name
User defined row name.
Last Calibrated
(Read only) Date and time that last calibration
was successfully executed.
Applied
(Read only)
“Yes”
if successfully applied
, “---“
if not yet
calibrated, and
“No”
if not applied.
Ext Mixer
(Read only)
Applies only if
Input = Ext Mixer
Type
Magnitude:
Select for correcting magnitude
only (no phase corrections). Use for swept
measurements.
Complex:
Select for correcting magnitude and
phase. Use for correcting IF responses for IQ
demodulation measurements.
Complex
is only available if RCal Option
CPX is installed.