U3050BH11
Electrostatic Discharge Protection
Electrostatic Discharge Protection
Protection against electrostatic discharge (ESD) is essential while removing assemblies from
or connecting cables to the network analyzer. Static electricity can build up on your body and
can easily damage sensitive internal circuit elements when discharged. Static discharges too
small to be felt can cause permanent damage. To prevent damage to the instrument:
•
always
have a grounded, conductive table mat (9300-0797) in front of your test equipment.
•
always
wear a grounded wrist strap (9300-1367) with grounding cord (9300-0980),
connected to a grounded conductive table mat, having a 1 MΩ resistor in series with it, when
handling components and assemblies or when making connections.
•
always
wear a heel strap (9300-1126) when working in an area with a conductive floor. If you
are uncertain about the conductivity of your floor, wear a heel strap.
•
always
ground yourself before you clean, inspect, or make a connection to a
static-sensitive device or test port. You can, for example, grasp the grounded outer shell of the
test port or cable connector briefly.
•
always
ground the center conductor of a test cable before making a connection to the
analyzer test port or other static-sensitive device. This can be done as follows:
1. Connect a short (from your calibration kit) to one end of the cable to short the center
conductor to the outer conductor.
2. While wearing a grounded wrist strap, grasp the outer shell of the cable connector.
3. Connect the other end of the cable to the test port and remove the short from the cable.
Figure
30
ESD Protection Setup
42
Keysight U3050BH11 User's and Service Guide
Summary of Contents for U3050BH11
Page 4: ...Contents 2...
Page 5: ...U3050BH11 Keysight U3050BH11 User s and Service Guide 1...
Page 21: ...U3050BH11 Operational Check Figure 13 Keysight U3050BH11 User s and Service Guide 17...
Page 23: ...U3050BH11 Operational Check Figure 15 Keysight U3050BH11 User s and Service Guide 19...
Page 25: ...U3050BH11 Operational Check Figure 17 Keysight U3050BH11 User s and Service Guide 21...
Page 27: ...U3050BH11 Operational Check Figure 19 Keysight U3050BH11 User s and Service Guide 23...
Page 29: ...U3050BH11 Operational Check Figure 21 Keysight U3050BH11 User s and Service Guide 25...
Page 31: ...U3050BH11 Operational Check Figure 23 Keysight U3050BH11 User s and Service Guide 27...
Page 33: ...U3050BH11 Operational Check Figure 25 Keysight U3050BH11 User s and Service Guide 29...