Troubleshooting the Test Set
-
Troubleshooting the Test Set
If the test set is not operating properly, use the following procedures to aid in
isolating the problem.
To request service, please contact your local service center. In the US, call
800-829-4444. For a listing of service centers worldwide, please visit us at
www.keysight.com/find/service.
No operator serviceable parts inside. Refer servicing to qualified
personnel. To prevent electrical shock do not remove covers.
Non-RF Failures
Non RF failures are; Power Supply, Fans, Control Lines, Status LEDs
or Interface not operating. Refer to
.
1. Verify that the front panel Power Switch is operational.The front panel
Active LED will be off unless the test set interface cable is connected and
the test set is addressed by the network analyzer.
2. Turn on the Standby Switch.
a. The rear panel Deck and Internal power supply fans should be
operational. If not continue with
(remove AC power from the instrument).” on page 74.
3. AC Line voltage checks (remove AC power from the instrument).
a. Ensure the proper AC Line voltage is present at the instrument line
cord.
b. Remove the AC power cord from the instrument. Verify the
instrument AC Line Module fuse (2119-0709) and replace if
necessary. Refer to
c. Verify the Internal AC lines fuse (2110-0342). Remove the
instrument bottom cover. Near the rear panel are two fuse holders,
verify the fuses and replace if necessary. See
The opening of covers or removal of parts is likely to expose the user to
dangerous voltages. Disconnect the instrument from all voltage
sources before being opened.