Keysight U3042AM20 User’s and Service Guide
9
U3042AM20
General Specifications
Refer to your analyzer's specifications to determine the maximum input
power levels for the analyzer's access and test ports, or to optimize the power
levels in the receivers.
Damage and maximum levels are not necessarily the optimum level.
Typical Reflection Tracking
Performance for the test set are typical. System performance for the analyzer
and test set are only characteristic and are intended as non-warranted
information.
Typical performance are based on 1 to 2 units performance, as shown in
Table 5
Typical Reflection Tracking PNA-X
Frequency
Ports 1 to 4
Ports 5 to 24
10 MHz to 50 MHz
1
1. Response from 10 MHz to 500 MHz is normal due to the
PNA-X couplers in comparison to the Test Set bridges. The
bridges have more gain in the coupled RF path.
−
0.5
> 20
50 MHz to 500 MHz
1
−
2
> 0
500 MHz to 3.2 GHz
−
4
−
5
3.2 GHz to 10 GHz
−
7
−
7
10 GHz to 16 GHz
−
8
−
8
16 GHz to 20 GHz
−
10
−
10
20 GHz to 24 GHz
−
12
−
12
24 GHz to 26.5 GHz
−
13
−
13
Summary of Contents for U3042AM20
Page 7: ...Keysight U3042AM20 User s and Service Guide 1 U3042AM20 ...
Page 105: ......