1-Port Calibration and Verification Procedure
1.
Connect the ECal or the mechanical cal kit to Port 1 or the port you are testing
on the PNA-X
.
Torque to 8
in-lb. For more information, refer to the Help menu (see Figure 48 below).
2.
Perform a 1-Port Calibration on Port 1. On the PNA-X, select Response > Cal > Start Cal >
Calibration Wizard.
a.
If using a mechanical cal kit, select SmartCal (Guided Calibration) > Next.
b.
If using an ECal module, connect the ECal to a PNA-X USB port. Select Use Electronic
Calibration (ECal) > Next.
3.
Continue following the Cal Wizard prompts. In the Select Calibration Ports and ECal Module
dialog box click Clear All, then select Port 1 > Next > Measure.
Figure 4
8
1-Port Calibration
4. Continue to follow the prompts until the Calibration Completed dialog box appears.
5. At the Calibration Completed prompt, select Save As User Calset, type the name 999.1.
Overwrite the Cal Set if it already exists and Save.
If you do not have a key board, select
Save As User Calset
>
Edit Name
and
save as
999.x
.
X is the port number you are calibrating. See Figure 4
9
below
. Use the numeric keypad on the PNA-X front panel to enter “999.1.”
Figure 4
9
Calibration Complete
Help
67
Keysight U3024AH10 User's and Service Guide
Summary of Contents for U3024AH10
Page 5: ...U3024AH10...
Page 52: ...Figure 38 Example 1 Test Set Signal Path 47 Keysight U3024AH10 User s and Service Guide...
Page 82: ...Figure 57 24 Port System Configuration 77 Keysight U3024AH10 User s and Service Guide...
Page 83: ...Figure 58 2 Way High Power Configuration 1 Watt Keysight U3024AH10 User s and Service Guide 78...
Page 105: ......