N5242A/B RF Interface Cable Connections
N5242A/B RF Interface Cable Connections
illustrates the cable configuration of the test set to the analyzer.
Cable Connection Procedure
1. Remove the SOURCE OUT to CPLR THRU and RCVR IN to CPLR ARM jumpers (x8) on the
analyzer. The RCVR R1 - R4 to SOURCE OUT reference loop jumpers (x4) remain on the front
panel.
2. Connect the RF interconnect cables from the PNA or PNA-X to the test set in the order listed.
As you are connecting each cable, torque to 8 in-lb. Refer to Table 6 below and
Over-torque will cause damage to the test set and may cause
connectors to spin or become loose.
Refer to
“Interconnect Cable Verification” on page 22
.
Table
6
PNA or
PNA-X RF Interface Cable Connection
(U3021-60077)
Numeric Order
RF Cables
From PNA-X
To Test Set
1
U3042-20122
Port 1 CPLR THRU
Port 3-14 CPLR THRU
2
U3042-20120
Port 1 SOURCE OUT
Port 3-14 SOURCE IN
3
U3042-20119
Port 1 CPLR ARM
Port 3-14 CPLR ARM
4
U3042-20121
Port 1 RCVR A IN
Port 3-14 RCVR OUT
5
U3042-20126
Port 2 CPLR THRU
Port 15-18 CPLR THRU
6
U3042-20128
Port 2 SOURCE OUT
Port 15-18 SOURCE IN
7
U3042-20125
Port 2 RCVR B IN
Port 15-18 RCVR OUT
8
U3042-20127
Port 2 CPLR ARM
Port 15-18 CPLR ARM
20
Keysight U3022AM42 User's and Service Guide