![Keysight Technologies U3022AM06 User'S And Service Manual Download Page 13](http://html1.mh-extra.com/html/keysight-technologies/u3022am06/u3022am06_users-and-service-manual_1962380013.webp)
U3022AM06
Available Options
Available Options
Test Set Option
s
The U3022AM06 three available options:
Refer to
“System Block Diagram” on page 7
Accessory Options
Front handle and rack flange accessories:
•
Option 1CM - Rack-Mount without Handles Kit (1CM114A)
•
Option 1CN - Front Handle Kit (1CN103A)
•
Option 1CP - Rack-Mount with Front Handle Kit (1CP106A)
Measurement Application Notes
Low Loss DUT Measurements
Low loss trace ripple is a product of switch repeatability and calibration
technique. Low insertion loss measurements exhibiting a 0.1 dB uncertainty
ripple residing on the measurement response is not uncommon.
Switch Configuration Command Speed
When writing address and data values to the test set directly, include a 15 ms
wait before issuing another address and data set. The delay of 15 ms is required
for the test set internal switches to settle. Failure to add the delay could cause
the test set internal switches to stick or cause intermittent failures.
•
Option 700 Standard, 300 kHz to 26.5 GHz frequency range
•
Option 129 External Noise source path for switch ports 1, 3, 5, and 7
for information on the Option 129 feature and control.