U3022AH10
Controlling the System with N5242A
/B
1.
On the PNA-X select STIMULUS > Sweep > Sweep Setup.
2.
Select Stepped Sweep.
3.
Set the Dwell Time to 5 μs > OK.
Figure
3
1
Sweep Setup
, N52xxA
Sweep Setup for Multiport and Standalone PNA-X Modes
When the Test Set is connected to the PNA-X, it is recommended that the analyzer’s Sweep
Setup be configured to Stepped Sweep before calibrating. This is slower than the Analog
Sweep, but is more accurate due to the extra electrical length of the Test Set and test port
cables. Only Stepped Sweep is available on all PNA-X models.
PNA-X
N52xxA
Models
1.
Select [Sweep] > Sweep Setup > Timing.
2.
Select
Timing tab
.
3.
Se
lect Sweep Mode: Stepped.
4.
Deselect Time: Auto Sweep Time.
5.
Set Dwell Time > Apply > OK.
Figure
3
2
Sweep Setup
, N52xxB
PNA-X
N52xxB
Models
U3022AH10 User's and Service Guide
39