U3022AH10
Controlling the System with N5242A
/B
Controlling the System with N5242A
/B
This section will describe how to setup and operate the U3022AH10 Multiport Test Set with the
N5242A
/B
2-Port or 4-Port PNA-X Network Analyzer.
The U3022AH10 Multiport Test Set is considered a “slave” instrument. The N5242A
/B
must
be used to control the Test Set. There are three methods to control the Test Set. Multiport mode
is recommended due to calibration and ease of use.
•
PNA-X Multiport Mode Option 551 requires firmware revision ≥ A.07.50.13. From the PNA-X
drop-down menu select Help > About Network Analyzer to see the options installed at the
installed firmware revision.
The following Test Set file must be installed into network analyzer file
directory: c:/program files/keysight/networkanalyzer. (see test set files link)
2-Port
PNA-X requires Test Set file:
u3022ae10_pnax_p2.tsx
4-Port PNA-X requires Test Set file:
u3022ae10_pnax_p4.tsx
Option 129 Test Set requires Test Set file:
u3022bh10_pnax_p
4
.tsx
•
PNA-X
Interface
Control
•
PNA-X GPIB Command Processor
Visit our website for firmware revision and downloads.
http://na.support.keysight.com/pna/firmware/firmware.html
http://na.support.keysight.com/multiport
The internal firmware of the
Keysight
N5242A
/B
has not been modified for
this Test Set option. Power levels may differ from those indicated on the
PNA-X when the Test Set is connected.
PNA-X Multiport Mode with Option 551
The PNA-X Multiport Mode selects the Test Set file that will enable the PNA-X to control the
Test Set. The PNA-X Multiport Mode allows you to complete a N-Port calibration using the Cal
Wizard application in the PNA-X. Refer to the PNA-X Help system for more information.
Typeface Key Conventions
The
following are key conventions used in this document:
- [HARDKEYS] are labeled front panel keys
- SOFTKEYS are indicated on the instrument display
-
(Italicized in parenthesis)
are menu paths for the 'B' Model analyzer
U3022AH10 User's and Service Guide
31