6 Verify Operation
6. 3. 3 Verify X-Apps Controls S9106AC-42D High IF Ports at 10 GHz
This process verifies operation of the
RF Transceiver 7 - 15 GHz
,
RF Out
➃
and
RF In
➄
located on the Rugged Front Panel;
these connectors may be referred to as: "High IF" or "DUT IF".
1. On RF Transceiver 1, connect an RF loop-back cable from
RF Out
➃
to
RF In
➄
located on the Rugged Front Panel.
This loop-back cable is not included with the
S9106AC-42D 5G FR1 Multi-port Transceiver.
Use a cable with two adapters (or equivalent) such as:
one 5063-1530 RF Cable, SMA (m) to SMA (m), 152.4 mm (6.0 in) and
two 1250-1250 Adapters, SMA (f) to Type-N (m)
72