Supplemental Characteristics - RP793xA, RP794xA
Supplemental characteristics are not warranted but are descriptions of performance determined
either by design or by type testing. Supplemental characteristics are typical unless otherwise noted.
Characteristic
RP7931A,
RP7941A
RP7932A,
RP7942A
RP7933A,
RP7943A
RP7935A,
RP7945A
RP7936A,
RP7946A
Minimum compliance limits
Voltage priority:
Current priority1:
±400 mA
0 and 20 mV
±125 mA
0 and 80 mV
±800 mA
0 and 20 mV
±250 mA
0 and 80 mV
±125 mA
0 and 160 mV
Voltage programming
Range:
Resolution:
0.02 to 20.4 V
191 µV
0.08 to 81.6 V
800 µV
0.02 to 20.4 V
191 µV
0.08 to 81.6 V
800 µV
0.16V to 163.2V
1.6 mV
Current programming
Range:
Resolution:
-408 A to 408 A
7.7 mA
-127.5 A to 127.5 A
2.5 mA
-816 A to 816 A
15.5 mA
-255 A to 255 A
5 mA
-127.5A to 127.5A
2.5 mA
Resistance programming
Range:
Resolution:
Accuracy:
0 to 0.098
Ω
0.8 µ
Ω
0.05% + 4 µ
Ω
0 to 1.25
Ω
9.8 µ
Ω
0.05% + 32 µ
Ω
0 to 0.049
Ω
0.4 µ
Ω
0.05% + 2 µ
Ω
0 to 0.625
Ω
4.8 µ
Ω
0.05% + 16 µ
Ω
0 to 2.5
Ω
19.6 µ
Ω
0.05% + 50 µ
Ω
Additional V-programming offset/ Ohm:
50 mV
15 mV
100 mV
30 mV
15 mV
Measurement range
Voltage:
Current:
-25 V to 25 V
-1014 A to 1014 A
-100 V to 100 V
-315 A to 315 A
-25 V to 25 V
-2028 A to 2028 A
-100 V to 100 V
-630 A to 630 A
-200 V to 200 V
-315 A to 315 A
Measurement digitization noise2
Voltage:
Current:
4 mV
400 mA
20 mV
200 mA
4 mV
800 mA
20 mV
400 mA
40 mV
200 mA
Programming & measurement
tempCo3
Voltage:
Current:
0.0015% +80 µV/°C
0.0075% +1.2mA/°C
0.0015% +300µV/°C
0.0025% +400µA/°C
0.0015% +80 µV/°C
0.0075% +2.3mA/°C
0.0015% +300µV/°C
0.0025% +800µA/°C
0.0015%+600µV/°C
0.0025%+400µA/°C
Output current ripple & noise (CC rms):
200 mA
70 mA
200 mA
70 mA
50 mA
Common mode current
CC rms:
CC peak-to-peak:
2 mA
20 mA
2.5 mA
24 mA
2 mA
20 mA
2.5 mA
24 mA
3.5 mA
30 mA
Overvoltage protection4
Maximum setting:
Response time:
Accuracy:
24 V
< 30 µs
0.02% + 2 mV
96 V
< 30 µs
0.02% + 8 mV
24 V
< 30 µs
0.02% + 2 mV
96 V
< 30 µs
0.02% + 8 mV
192 V
< 30 µs
0.02% + 16 mV
Voltage Programming Speed, no load5
Rise/fall time 10% to 90% of step:
Settling time to 0.1% of step:
80 µs
810 µs
75 µs
480 µs
80 µs
810 µs
75 µs
480 µs
75 µs
550 µs
Voltage Programming Speed, full load6
Rise/fall time 10% to 90% of step:
Settling time to 0.1% of step:
140 µs
4.2 ms
130 µs
1.35 ms
140 µs
4.2 ms
130 µs
1.35 ms
170 µs
1.35 ms
Current Up/down Programming Speed7
Rise/fall time 10% to 90% of step:
Settling time to 0.1% of step::
300 µs
960 µs
180 µs
500 µs
300 µs
960 µs
180 µs
500 µs
190 µs
550 µs
1 Quick Reference
36
Keysight RP7900 Series Operating and Service Guide
Summary of Contents for RP7900 Series
Page 1: ...Keysight Regenerative Power System RP7900 Series Operating and Service Guide ...
Page 2: ......
Page 100: ......
Page 262: ......
Page 314: ......