INSTrument Subsystem
Instrument commands program he Master/Slave function of the instrument. Master Slave operation is
used when connecting a number of instruments in parallel to create a system with higher total current
and, hence, higher power.
INSTrument:GROup:FUNCtion MASTer|SLAVe|NONe
INSTrument:GROup:FUNCtion?
Set the function of a unit in a master/slave configuration. This setting is saved in non-volatile memory.
MASTer - Configures the instrument as the master unit in a master/slave group.
SLAVe - Configures the instrument as a slave unit in a master/slave group.
NONE - disables the master/slave function. The unit operates independently.
Parameter
Typical Return
MASTer | SLAVe | NONE
MAST, SLAV, or NONE
Configures the instrument as the master INST:GRO:FUNC MAST
INSTrument:GROup:MASTer:CONNect[:STATe] [,0|OFF|1|ON]
Instructs the master unit to connect to the previously discovered slave units. If the slave configuration
on the bus matches the discovered one, the master will start normal operation. Otherwise the
command will fail with an error, and all units will operate independently.
Parameter
Typical Return
optional 0|OFF|1|ON
0 or 1
Connect the master to the slave units INST:GRO:MAST:CONN
INSTrument:GROup:MASTer:CONNect:DELay <
value
>|MIN|MAX
INSTrument:GROup:MASTer:CONNect:DELay? [MIN|MAX]
Set the delay after power-on before the master units attempts to connect to slave units. This only
applies if the connect mode is set to AUTO. This setting is saved in-volatile memory.
Parameter
Typical Return
0 to 120 seconds
0
Configures the connection delay for 10 seconds INST:GRO:MAST:CONN:DEL 10
5 SCPI Programming Reference
202
Keysight RP7900 Series Operating and Service Guide