Keysight Double-Pulse Test Rack and Safety Enclosure Operation Guide
17
Electrical and Environmental Requirements
Introduction
Electrical and Environmental Requirements
Rack and Safety Enclosure Dimensions and Weight
Input Voltage
Operating
Temperature
Operating Humidity
Operating
Environment
200-240 VAC,
10 %,
Maximum Power rating 13A,
Typical 600 VA, 50/60 Hz
20 °C to +30 °C
50% to 80% RH,
Non-condensing
Indoor location only
The Double-Pulse Test Rack and Safety Enclosure is designed to
operate on 200 to 240 VAC single phase. Do not attempt to
operate the DPT Rack and Safety Enclosure on 120 VAC.
Avoid overloading an electrical circuit. Ensure your AC line
matches the AC Input requirements. If you need to power
down the equipment in an emergency, make sure that you
have clear and quick access to the primary disconnect.
Refer to “Mains Disconnect” below.
Weight
Height
Depth
Width
Shipping Crate (with DPT Rack and
Safety Enclosure)
300 kg
(660 lbs)
167 cm
(65.75 in)
92 cm
(36.25 in)
120 cm
(47.25 in)
Rack and Safety Enclosure alone
(no test equipment installed)
150kg
(330.7lbs)
160 cm
(63 in.)
70 cm
(27.6 in.)
100 cm
(39.4 in.)
Rack and Safety Enclosure
(all test equipment and fixtures
installed)
162.5 kg
(358 lbs