82 N9030B PXA Signal Analyzer Service Guide
Boot Up and Initialization Troubleshooting
Signal Level Verification
Signal Level Problem with
Input Frequencies > 3.6 GHz
Measure the 4.8 GHz RF calibrator signal level by pressing
Input/Output
,
RF
Calibrator
,
4.8 GHz
. Now press
Freq
,
4.8 GHz
,
SPAN
,
1 MHz
,
Peak Search
. If
the analyzer is functioning correctly in high band, the 4.8 GHz calibrator level
should be
−
28 dBm
±
2 dB. See the figure below.
Is the signal level correct?
If yes:
The analyzer high band path is functioning properly at least at 4.8 GHz.
If no:
Press
Amplitude
,
Y-Scale
,
Preselector Center
to manually center the
preselector. Refer to
Chapter 4, “RF Section Troubleshooting
in this manual.
It is possible that other input frequencies > 3.6 GHz could have an
amplitude problem even though the 4.8 GHz calibrator is within tolerance.
Using the internal RF calibrator provides a quick check of the high band
path.