N9030B PXA Signal Analyzer Service Guide 81
Boot Up and Initialization Troubleshooting
Signal Level Verification
Signal Level Verification
Signal Level Problem with
Input Frequencies < 3.6 GHz
Measure the 50 MHz RF calibrator signal level by pressing
Input/Output
,
RF
Calibrator
,
50 MHz
. Now press
Freq
,
50 MHz
,
SPAN
,
1 MHz
,
Peak Search
. If
the analyzer is functioning correctly in low band, the 50 MHz calibrator level
should be
−
25 dBm
±
2 dB. See the figure below.
Is the signal level correct?
If yes:
The analyzer low band path is functioning properly at least at 50 MHz.
If no:
Chapter 4, “RF Section Troubleshooting (RF/Microwave
It is possible that other input frequencies < 3.6 GHz could have an
amplitude problem even though the 50 MHz calibrator is within tolerance.
Using the internal RF calibrator provides a quick check of the low band
path.