N9030B PXA Signal Analyzer Service Guide 593
Functional Tests
Frequency Response (Flatness), Preamp On
Frequency Response (Flatness), Preamp On
Test Limits
The frequency response test, with preamplifier on, measures the signal
analyzer’s amplitude error as a function of the tuned frequency. Measurements
are made ranging from 50 MHz to the maximum frequency range of the
preamp. The signal source amplitude is measured with a power meter to
eliminate error due to source flatness. The measured value is normalized to 50
MHz.
Frequency Range
Limit Relative to 50 MHz
100 kHz to 3.6 GHz
±2.0 dB
> 3.6 GHz to 8.4 GHz
±3.0 dB
> 8.4 GHz to 13.6 GHz
±3.5 dB
> 13.6 GHz to 26.5 GHz
±4.0 dB
> 26.5 GHz to 34.5 GHz
±4.0 dB
> 34.5 GHz to 50 GHz
±5.0 dB
Item
Critical Specifications
(for this test)
Recommended
Keysight Model
Adapter
Type N (m) to 3.5 mm (m)
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
1250-1743
Adapter
3.5 mm (f) to 3.5 mm (f)
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.1:1
83059B
Cable, 3.5 mm, 1 meter
(Opt P03, P08, P13, P26)
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.4:1
11500E
Cable, 2.4 mm, 1 meter
(Opt P43, P44, P50)
Frequency: 10 MHz to 50 GHz
VSWR: < 1.55:1
8120-6164
Cables, BNC
120 cm
Frequency: 10 MHz
10503A
Power Meter
Compatible with power sensor
E4418B
Attenuator, 30 dB Fixed
Frequency: 50 MHz
Accuracy: ±0.05 dB
11708A
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...