N9030B PXA Signal Analyzer Service Guide 57
Boot Up and Initialization Troubleshooting
Potential Problems During Boot Process
Potential Problems During Boot Process
This section describes potential problems that may occur if there is an internal
hardware issue that prohibits the instrument from completing a full boot up to
the spectrum analyzer application.
Yellow Standby LED Does Not Illuminate
Control of the yellow front panel Standby LED comes from the A4 Processor
board assembly. This signal is routed through the A7 Midplane board and is
then buffered on the A8 Motherboard before being sent to the A1A2 Front
Panel Interface board through W1. Of course, the power for this all originates
with the A6 Power Supply Assembly. When the Standby LED does not come on
it could be due to any one of these assemblies. This procedure will help to
determine which one is the cause.
If the instrument turns on and operates properly but the yellow Standby LED
does not work then all that will need to be done is to trace where the control
signal for the LED is being lost using the routing information in the preceding
paragraph.
1.
The Standby LED will only turn on when the instrument is connected to an
AC power source that has a voltage level and frequency of that specified
for the instrument. Before proceeding verify that these requirements are
being met. Refer to the instrument rear panel for these requirements.
2.
Remove the AC power cord and then remove the instrument cover. Refer
Chapter 16, “Assembly Replacement Procedures”, on page 429
in this
manual.
3.
, verify the +5.1VSB LED on the A7 Midplane board
is on (green).
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...