440 N9030B PXA Signal Analyzer Service Guide
Assembly Replacement Procedures
RF Area (Options 503, 508, 513, 526)
Attenuators
Removal
1.
. To remove Attenuator A
(1)
or Attenuator B
(2)
,
remove the semi-rigid cables W2, W3, W4, or W6 attached to the
attenuator using the 5/16 inch wrench.
2.
Remove the ribbon cable attached to the attenuator.
3.
Remove the two screws
(3)
or
(4)
(
) for each attenuator
bracket to remove from the chassis using the T-10 driver.
4.
Remove the attenuator from the bracket by removing the two screws
(
).
Summary of Contents for N9030B