Specifications and Characteristics
2
N8480 Series Power Sensors Operating and Service Guide
39
Calibration Factor Uncertainty
Table 2-8
Calibration factor uncertainty at
25 ºC ± 3 ºC
Frequency
25 ºC ± 3 ºC
N8481A
N8481B
N8481H
N8482A
N8482B
N8482H
100 kHz to 10 MHz
-
-
-
0.91
1.48
0.89
10 MHz to 30 MHz
0.82
1.42
0.77
0.78
1.43
0.79
30 MHz to 500 MHz
0.77
1.48
0.89
0.77
1.49
0.89
500 MHz to 1.2 GHz
0.78
1.48
0.89
0.78
1.49
0.89
1.2 GHz to 6 GHz
0.91
1.58
1.06
0.89
1.56
1.02
6 GHz to 14 GHz
1.26
1.77
1.46
-
-
-
14 GHz to 18 GHz
1.59
1.92
1.73
-
-
-
Table 2-8
Calibration factor uncertainty at
25 ºC ± 3 ºC (continued)
Frequency
25 ºC ± 3 ºC
N8485A
N8487A
N8486AR
N8486AQ
100 kHz to 10 MHz
-
-
-
-
10 MHz to 30 MHz
0.82
-
-
-
30 MHz to 500 MHz
1.24
1.33
-
-
500 MHz to 1.2 GHz
1.26
1.35
-
-
1.2 GHz to 6 GHz
1.35
1.41
-
-
6 GHz to 14 GHz
1.66
1.61
-
-
14 GHz to 18 GHz
1.83
1.73
-
-
18 GHz to 26.5 GHz
2.67
2.26
-
-
26.5 GHz to 33 GHz
3.32
2.58
2.68
-
33 GHz to 34 GHz
-
2.80
3.19
3.14
34 GHz to 35 GHz
-
2.80
3.19
3.40
35 GHz to 40 GHz
-
2.80
3.19
3.14
40 GHz to 45 GHz
-
3.66
-
3.19
45 GHz to 50 GHz
-
4.23
-
3.26