![Keysight Technologies N8262A Programming Manual Download Page 441](http://html.mh-extra.com/html/keysight-technologies/n8262a/n8262a_programming-manual_1962053441.webp)
STATus Subsystem
11
Keysight N8262A Programming Guide
441
STATus:QUEStionable:POWer[:SUMMary]
The questionable power summary register set contain bits that indicate the
quality of the power data being acquired by the power meter.
The following bits in these registers shall be used by the power meter:
Bit 1 is set when any of the following errors occur:
– Error –231, “Data questionable;Input Overload”
– Error –231, “Data questionable;Input Overload ChA”
Bit 2 is set when the following error occurs:
– Error –231, “Data questionable;Input Overload ChB”
Bits 3 is set when the following error occurs:
– Error –230, “Data corrupt or stale”
– Error –231, “Data questionable;Upper window log error”
Bit 4 is set when the following error occurs:
– Error –230, “Data corrupt or stale”
– Error –231, “Data questionable;Lower window log error”
Bit 5 is set when the following condition occurs:
Bit Number
Decimal Weight
Definition
0
-
Not used
1
2
Channel A Power
2
4
Channel B Power
3
8
Upper Window Power
4
16
Lower Window Power
5
32
Channel A Please Zero
6
64
Channel B Please Zero
7
128
Upper Window Lower Measurement Power
8
256
Lower Window Lower Measurement Power
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...