STATus Subsystem
11
Keysight N8262A Programming Guide
439
STATus:QUEStionable
The questionable register set contains bits that indicate the quality of various
aspects of signals processed by the power meter.
The following bits in these registers are used by the power meter:
Bit 3 is set by the logical OR outputs of the
STATus:QUEStionable:POWer:SUMMary
register set.
Bit 8 is set by the logical OR outputs of the
STATus:QUEStionable:CALibration:SUMMary
register set.
Bit 9 is set if power-on self-test fails, and cleared if it passes.
Syntax
Bit Number
Decimal Weight
Definition
0 to 2
-
Not used
3
8
POWer Summary
4 to 7
-
Not used
8
256
CALibration Summary
9
512
Power On Self Test
10 to 15
-
Not Used (bit 15 always 0)