SENSe Subsystem
10
Keysight N8262A Programming Guide
369
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2[:SELect] <string>
This command enters the name of the sensor calibration table or frequency
dependent offset table which is to be used. The
CSET1
command selects the
sensor calibration table and the
CSET2
command selects the frequency dependent
offset table. The calibration factor is interpolated from the table using the setting
for
[SENSe[1]]|SENSe2:FREQuency
.
Syntax
Parameters
NOTE
If
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe
is set to
OFF
, the
selected sensor calibration table or frequency offset table is not being used.
Item
Description/Default
Range of Values
string
String data representing a sensor calibration table, or
frequency dependent offset table name.
Any existing table name
(Existing table names
can be listed using
MEMory:CATalog:TABle?
).
SENS
1
:CSET
CORR
SENS2
:
?
:SEL
string
Space
1
:CSET2
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...