OUTPut Subsystem
8
Keysight N8262A Programming Guide
317
OUTPut:RECorder[1]|2:FEED <data_handle>
This command specifies which measurement is sent to the recorder output
specified by the numeric value following
RECorder
,
RECorder1
and
RECorder2
.
Syntax
Parameters
Example
Reset Condition
On reset,
data_handle
is set to its previous value.
OUTP
:FEED
:REC
Space
?
1
2
data_handle
Item
Description/Default
Range of Values
data_handle
The CALC block specifying the measurement to be sent to
the recorder output.
“CALC1”
or
“CALC”
“CALC2”
“CALC3”
“CALC4”
OUTP:REC2:FEED “CALC1”
This command sends the CALC1
measurement to recorder output 2.
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...