MEMory Subsystem
7
Keysight N8262A Programming Guide
287
MEMory:CATalog:STATe?
This command is used to list the save/recall registers.
The power meter returns the data in the form of two numeric parameters and as
many strings as there are save/recall registers.
<numeric_value>,<numeric_value>{,<string>}
– The first numeric parameter indicates the amount of memory, in bytes, used
for the storage of registers.
– The second parameter indicates the memory, in bytes, available for the
storage of registers.
– Each string parameter returned indicates the name, type and size of a save/
recall register:
–
<string>,<type>,<size>
–
<string>
indicates the name of the save/recall register.
–
<type>
indicates
STAT
for save/recall register.
–
<size>
indicates the size of the save/recall register in bytes.
For example, a sample of a response may look like:
0,16190,"State0,STAT,0","State1,STAT,0" .........
Syntax
Example
MEM:CAT:STAT?
This command queries the list of save/recall
registers.
MEM
:CAT
:STAT
?
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...