CALibration Subsystem
234
Keysight N8262A Programming Guide
CALibration[1]|2:RCFactor <numeric_value>
This command is used with 8480 Series power sensors to set the reference
calibration factor of the specified channel. Reference calibration factors can also
be set using sensor calibration tables. The power meter uses the most recently set
reference calibration factor.