CALibrate:STATe 0|OFF|1|ON [,<password>]
CALibrate:STATe?
Enables or disables calibration mode. Calibration mode must be enabled for the instrument to accept
any calibration commands. The first parameter specifies the state. The second optional parameter is
the password.
Parameter
Typical Return
0|OFF|1|ON, *RST OFF
0 or 1
<
password
> a numeric value up to 15 digits
(none)
Disable calibration: CAL:STAT OFF
Enable calibration: CAL:STAT ON [,value]
A password is required if <
password
> has been set to a non-zero value.
CALibrate:VOLTage[:LEVel] <
value
>
Calibrates the local voltage programming and measurement. The value selects the range to
calibrate.
Parameter
Typical Return
The maximum voltage of the output range.
(none)
Calibrates the voltage of the 20 V range: CAL:VOLT 20
5 SCPI Programming Reference
158
Keysight N69100 Series Operating and Service Guide