![Keysight Technologies N69100 Series Operating And Service Manual Download Page 122](http://html1.mh-extra.com/html/keysight-technologies/n69100-series/n69100-series_operating-and-service-manual_1962218122.webp)
Specify the Arb type and dwell
To specify the Arb type and dwell:
Front Panel Menu Reference
SCPI Command
Select
Transient\Arb\Config
.
In the dropdown list, pick either a voltage or a
current Arb. Then press
Select
.
Enter a dwell value in the Dwell field. Then press
Select
.
To specify a voltage or current Arb:
ARB:FUNC:TYPE VOLT
ARB:FUNC:TYPE CURR
To specify a dwell time of 1 millisecond:
ARB:VOLT:CDW:DWEL 0.001
ARB:CURR:CDW:DWEL 0.001
Configure the Arb
Note that you can use the numeric up/down arrows to view arb points in the front panel
Transient\Arb\Config dialog. You cannot
program
Arb data from the front panel. You must use the
SCPI ARB:CURRent:CDWell or ARB:VOLTage:CDWell commands to program the Arb data.
Front Panel Menu Reference
SCPI Command
Select
Transient\Arb\Config
.
If Arb points have been imported or programmed
using the SCPI command, the Points field displays
the number of points in the Arb.
View the amplitude of any Arb point by entering the
point number in the Point # field. The Level field
displays the amplitude.
To program 10 points in a current Arb:
ARB:CURR:CDW 1,2,2,3,4,4,3,2,2,1
To query the number of Arb points:
To query the Arb point values:
Specify how many times you want the Arb to repeat
Depending on your application, specify how many times you want the Arb to repeat. Sending the
INFinity parameter in the SCPI command makes the Arb repeat indefinitely. At reset, the Arb count is
set to 1.
Front Panel Menu Reference
SCPI Command
Select
Transient\Arb\Repeat
.
Enter the number of list repetitions (2) and press
Select.
To program the Arb to repeat twice:
Specify how you want the Arb to end
Specify the output state after the Arb has completed. There are two choices: the output returns to the
state it was in before the Arb started, or the output remains at the values of the last Arb point.
Front Panel Menu Reference
SCPI Command
Select
Transient\Arb\Terminate
.
Select either Return to Start, or Stop at Last Step
and press Select.
To return the output to the pre-Arb state:
To keep the output at the Arb end point:
4 Using the DC Power System
122
Keysight N69100 Series Operating and Service Guide