Voltage Programming Response
The following graphs show the output voltage programming response characteristic. These graphs
apply for small signals only, under no load condition.
Output Dynamic Response
The following graph shows the square wave amplitude thresholds versus frequency for each model.
Above the indicated amplitude threshold, sustained generation of a square wave may lead to
engagement of the excessive dynamic protection (
) function, which disables the output. EDP
protection may be engaged by programmed voltage changes, lists, Arbs, or load-induced voltage
swings.
Maximum Peak-to-Peak Programmable Voltage vs. Square Wave Frequency
1 Quick Reference
60
Keysight N6900/N7900 Series Operating and Service Guide