11
PROT
2048
Output is disabled by a watchdog timer protection
12
EDP
4096
Output is disabled by excessive output dynamic protection
13
SF
8192
A sense lead fault has been detected
14,15
not used
not used
0 is returned
The following table describes the Questionable2 assignments.
Bit
Bit Name
Decimal Value
Definition
0
UProt
1
Output is disabled by a user-defined protection signal
1
IPK+
2
Output is in positive peak current limit
2
IPK-
4
Output is in negative peak current limit
3
CSF
8
A current sharing fault has occurred
4-15
not used
not used
0 is returned
Standard Event Status Group
These registers are programmed by Common commands. The group consists of an Event and Enable
register. The Standard Event event register latches events relating to communication status. It is a
read-only register that is cleared when read. The Standard Event enable register functions similarly
to the enable registers of the Operation and Questionable status groups. Refer to
for
a description of each register. The following table describes the Standard Event Status register bit
assignments.
Bit
Bit Name
Decimal
Value
Definition
0
Operation Complete
1
All commands before and including *OPC have been executed.
1
not used
not used
0 is returned
2
Query Error
4
The instrument tried to read the output buffer but it was empty, a new command line was
received before a previous query has been read, or both the input and output buffers are
full.
3
Device-specific Error
8
A device-specific error, including a self-test error, calibration error or other device-spe-
cific error occurred.
4
Execution Error
16
An execution error occurred.
5
Command
32
A command syntax error occurred.
6
not used
not used
0 is returned
7
Power On
128
Power has been cycled since the last time the event register was read or cleared.
6 SCPI Programming Reference
287
Keysight N6900/N7900 Series Operating and Service Guide
Summary of Contents for N6900 Series
Page 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Page 2: ......
Page 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Page 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Page 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...