[SOURce:]CURRent:LIMit[:POSitive][:IMMediate][:AMPLitude] <
value
>|MIN|MAX
[SOURce:]CURRent:LIMit[:POSitive][:IMMediate][:AMPLitude]? [MIN|MAX]
[SOURce:]CURRent:LIMit:NEGative[:IMMediate][:AMPLitude] <
value
>|MIN|MAX
[SOURce:]CURRent:LIMit:NEGative[:IMMediate][:AMPLitude]? [MIN|MAX]
Sets the current limit when in voltage priority mode. Units are in amperes. The maximum value
depends on the current rating of the unit. The minimum value is the most negative value.
Parameter
Typical Return
Positive: 0 to 102% of rating, *RST 1.02% of rating
Negative: -10.2% of rating to 0, *RST -10.2% of rating
Negative with power dissipator: up to -102% of rating to 0
<+current limit>
<–current limit>
Sets the positive current limit to 2 A: CURR:LIM 2
Sets the negative current limit to -2 A: CURR:LIM:NEG -2
[SOURce:]CURRent:MODE FIXed|STEP|LIST|ARB
[SOURce:]CURRent:MODE?
Sets the transient mode. This determines what happens to the output current when the transient
system is initiated and triggered.
FIXed keeps the output current at its immediate value.
STEP steps the output to the triggered level when a trigger occurs.
LIST causes the output to follow the list values when a trigger occurs.
ARB causes the output to follow the arbitrary waveform values when a trigger occurs.
Parameter
Typical Return
FIXed|STEP|LIST|ARB, *RST FIXed
FIX, STEP, LIST, or ARB
Sets the current mode to Step: CURR:MODE STEP
[SOURce:]CURRent:PROTection:DELay[:TIME] <
value
>|MIN|MAX
[SOURce:]CURRent:PROTection:DELay[:TIME]? [MIN|MAX]
Sets the over-current protection delay. The over-current protection function will not be triggered
during the delay time. After the delay time has expired, the over-current protection function will be
active. This prevents momentary changes in output status from triggering the over-current protection
function. Values up to 255 milliseconds can be programmed, with a resolution of 1 millisecond.
Parameter
Typical Return
0 – 0.255, *RST 0.020 s
<delay value>
Sets the protection delay to 0.2 seconds: CURR:PROT:DEL 0.2
l
The operation of over-current protection is affected by the setting of the current protection delay
start event, which is specified by CURRent:PROTection:DELay:STARt.
6 SCPI Programming Reference
219
Keysight N6900/N7900 Series Operating and Service Guide
Summary of Contents for N6900 Series
Page 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Page 2: ......
Page 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Page 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Page 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...