CALibrate:CURRent:MEASure <
value
>
Calibrates the low range current measurement. The value selects the range to calibrate.
Parameter
Typical Return
The maximum current of the measurement range.
(none)
Calibrates the current of the 10 A measurement range:
CAL:CURR:MEAS 10
CALibrate:CURRent:SHARing
Calibrates the Imon signal for paralleled units.
Parameter
Typical Return
(none)
(none)
Calibrates the current sharing: CAL:CURR:SHAR
Calibrates the temperature coefficient.
Parameter
Typical Return
(none)
(none)
Calibrates the temperature coefficient: CAL:CURR:TC
CALibrate:DATA <
value
>
Enters the calibration value read by the external meter. You must first select a calibration level for the
value being entered. Data values are expressed in base units - either volts or amperes, depending on
which function is being calibrated.
Parameter
Typical Return
Numeric value
(none)
Specify calibration value 0.0237: CAL:DATA 2.37E-2
CALibrate:DATE <
"date"
>
CALibrate:DATE?
Enters the calibration date in nonvolatile memory. Enter any ASCII string up to 15 characters. The
query returns the date. Enclose string parameters in single or double quotes.
6 SCPI Programming Reference
215
Keysight N6900/N7900 Series Operating and Service Guide
Summary of Contents for N6900 Series
Page 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Page 2: ......
Page 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Page 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Page 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...