CALibrate Subsystem
Calibrate commands calibrate the instrument.
Read the
before calibrating. Improper calibration reduces
accuracy and reliability.
CALibrate
:COUNt?
Returns the number of times the unit has been calibrated.
:CURRent
[:LEVel] <
value
>
Calibrates the current programming and measurement.
:MEASure <
value
> Calibrates the low range current measurement.
:SHARing
Calibrates the Imon signal for paralleled units.
:TC
Calibrates the temperature coefficient.
:DATA <
value
>
Enters the calibration value read by the external meter.
:DATE <
"date"
>
Enters the calibration date in nonvolatile memory.
:LEVel P1|P2|P3
Advances to the next level in the calibration.
:PASSword <
value
>
Sets a numeric password to prevent unauthorized calibration.
:RESistance
:BOUT
Calibrates the bottom out resistance.
:SAVE
Saves the calibration constants in non-volatile memory.
:STATe 0|OFF|1|ON
Enables or disables calibration mode.
:VOLTage
[:LEVel] <
value
>
Calibrates the local voltage programming and measurement.
:CMRR
Calibrates the voltage common mode rejection ratio.
CALibrate:COUNt?
Returns the number of times the unit has been calibrated. The count is incremented whenever the
calibration (and date) is saved, the administration password is changed or reset, or the firmware is
updated.
Parameter
Typical Return
(none)
<count>
Return the calibration count: CAL:COUN?
CALibrate:CURRent[:LEVel] <
value
>
Calibrates the current programming and measurement. The value selects the range to calibrate.
Parameter
Typical Return
The maximum current of the output range.
(none)
Calibrates the current of the 10 A range: CAL:CURR 10
Keysight N6900/N7900 Series Operating and Service Guide
214
6 SCPI Programming Reference
Summary of Contents for N6900 Series
Page 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Page 2: ......
Page 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Page 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Page 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...