ABORt Subsystem
Abort commands cancel any triggered actions and returns the trigger system back to the Idle state.
Abort commands are also executed with the *RST command.
ABORt:ACQuire
ABORt:ELOG
ABORt:TRANsient
ABORt:ACQuire - Cancels any triggered measurements. It also resets the WTG-meas and MEAS-
active bits in the Operation Status registers.
ABORt:ELOG - Stops external data logging. It also resets the WTG-meas and MEAS-active bits in the
Operation Status registers.
ABORt:TRANsient - Cancels any triggered actions. It also resets the WTG-tran and TRAN-active bits in
the Operation Status registers. Note that this command does not turn off continuous triggers if
INITiate:CONTinuous:TRANsient ON has been programmed. In this case, the trigger system will
automatically re-initiate.
Parameter
Typical Return
(none)
(none)
Aborts the triggered measurement: ABOR:ACQ
Keysight N6900/N7900 Series Operating and Service Guide
210
6 SCPI Programming Reference
Summary of Contents for N6900 Series