An expression program data element related error occurred.
−261 Math error in expression
An expression program data element could not be executed due to a math error.
Query Errors (these errors set Standard Event Status register bit #2)
-400 Query Error
Generic error query
-410 Query INTERRUPTED
A condition causing an interrupted query error occurred.
-420 Query UNTERMINATED
A condition causing an unterminated query error occurred.
-430 Query DEADLOCKED
A condition causing a deadlocked query error occurred.
-440 Query UNTERMINATED after indefinite response
A query was received in the same program message after a query indicating an indefinite response
was executed.
Selftest Errors (these errors set Standard Event Status register bit #3)
202 Selftest Fail Aux Adc 0 expected <n1> to <n2>, measured <n3>, chan <n4>
Auxiliary ADC failed. n1 and n2 are the expected limits. n3 is the measured value. n4 is the channel
location of the failed module.
202 Selftest Fail DACs 0 expected <n1> to <n2>, measured <n3>, chan <n4>
Both voltage and current DACs are at zero. n1 and n2 are the expected limits. n3 is the measured
value. n4 is the channel location of the failed module.
202 Selftest Fail DACs 1 expected <n1> to <n2>, measured <n3>, chan <n4>
Voltage DAC is at zero; current DAC is at full scale. n1 and n2 are the expected limits. n3 is the
measured value. n4 is the channel location of the failed module.
202 Selftest Fail DACs 2 expected <n1> to <n2>, measured <n3>, chan <n4>
Voltage DAC is at full scale; current DAC is at zero. n1 and n2 are the expected limits. n3 is the
measured value. n4 is the channel location of the failed module.
202 Selftest Fail DACs 3 expected <n1> to <n2>, measured <n3>, chan <n4>
Both voltage and current DACs are at full scale. n1 and n2 are the expected limits. n3 is the measured
value. n4 is the channel location of the failed module.
7 SCPI Programming Reference
368
Keysight N6705C Operating and Service Guide
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......