Parameter
Typical Return
VOLTage|CURRent, *RST VOLTage
VOLT or CURR
Sets the preferred mode to CURRent: OUTP:PMODE CURR, (@1)
OUTPut[:STATe]:TMODe HIGHz|LOWz, (@<chanlist>)
OUTPut[:STATe]:PMODe? (@<chanlist>)
Specifies the output impedance mode at turn-off.
HIGHz - sets the output to a high impedance mode at turn-off. In high impedance mode, the output
relays are opened while the output remains at its set value.
LOWZ - sets the output to a low impedance mode at turn-off. In low impedance mode, the output is
first down-programmed to zero, after which the output relays are opened.
Parameter
Typical Return
HIGHz|LOWZ, *RST HIGHz
HIGH or LOWZ
Sets the turn-off mode to high impedance: OUTP:TMODE HIGH, (@1)
OUTPut:INHibit:MODE LATChing|LIVE|OFF
OUTPut:INHibit:MODE?
Sets the operating mode of the
digital pin. The inhibit function shuts down the output in
response to an external signal on the Inhibit input pin. The parameter is saved in non-volatile memory.
LATChing - a logic-true signal on the Inhibit input causes the output state to latch OFF. The output
remains disabled until the Inhibit input is returned to logic-false and the latched INH status bit is
cleared by sending OUTPut:PROTection:CLEar or a protection clear from the front panel.
LIVE - allows the enabled output to follow the state of the Inhibit input. When the Inhibit input is true,
the output is disabled. When the Inhibit input is false, the output is re-enabled.
OFF - The Inhibit input is ignored.
Parameter
Typical Return
LATChing|LIVE|OFF
LATC, LIVE, or OFF
Sets the Inhibit input to latching mode: OUTP:INH:MODE LATC
OUTPut:PON:STATe RST|RCL0
OUTPut:PON:STATe?
This determines whether the power-on state is set to the *RST state (RST) or the state stored in
memory location 0 (RCL0). Instrument states can be stored using the *SAV command. This parameter
is saved in non-volatile memory.
7 SCPI Programming Reference
310
Keysight N6705C Operating and Service Guide
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......